Meyer, Sebastian; Pinto, Marco; Parodi, Katia und Gianoli, Chiara (2021): The impact of path estimates in iterative ion CT reconstructions for clinical-like cases. In: Physics in Medicine and Biology, Bd. 66, Nr. 9, 95007

Volltext auf 'Open Access LMU' nicht verfügbar.

Abstract

Ion computed tomography (CT) promises to mitigate range uncertainties inherent in the conversion of x-ray Hounsfield units into ion relative stopping power (RSP) for ion beam therapy treatment planning. To improve accuracy and spatial resolution of ion CT by accounting for statistical multiple Coulomb scattering deflection of the ion trajectories from a straight line path (SLP), the most likely path (MLP) and the cubic spline path (CSP) have been proposed. In this work, we use FLUKA Monte Carlo simulations to investigate the impact of these path estimates in iterative tomographic reconstruction algorithms for proton, helium and carbon ions. To this end the ordered subset simultaneous algebraic reconstruction technique was used and coupled with a total variation superiorization (TVS). We evaluate the image quality and dose calculation accuracy in proton therapy treatment planning of cranial patient anatomies. CSP and MLP generally yielded nearly equal image quality with an average RSP relative error improvement over the SLP of 0.6%, 0.3% and 0.3% for proton, helium and carbon ion CT, respectively. Bone and low density materials have been identified as regions of largest enhancement in RSP accuracy. Nevertheless, only minor differences in dose calculation results were observed between the different models and relative range errors of better than 0.5% were obtained in all cases. Largest improvements were found for proton CT in complex scenarios with strong heterogeneities along the beam path. The additional TVS provided substantially reduced image noise, resulting in improved image quality in particular for soft tissue regions. Employing the CSP and MLP for iterative ion CT reconstructions enabled improved image quality over the SLP even in realistic and heterogeneous patient anatomy. However, only limited benefit in dose calculation accuracy was obtained even though an ideal detector system was simulated.

Dokument bearbeiten Dokument bearbeiten