Molina, Rafael; Gion, Massimo; Gressner, Axel; Troalen, Fredric; Auge, Jose Maria; Holdenrieder, Stefan; Zancan, Matelda; Wycislo, Matthias und Stieber, Petra (2008): Alternative antibody for the detection of CA15-3 antigen: a European multicenter study for the evaluation of the analytical and clinical performance of the Access (R) BR Monitor assay on the UniCel (R) Dxl 800 Immunoassay System. In: Clinical Chemistry and Laboratory Medicine, Bd. 46, Nr. 5: S. 612-622 [PDF, 199kB]

[thumbnail of cclm.2008.133.pdf]
Vorschau
Download (199kB)

Abstract

Background: Cancer antigen CA15-3 antigen is known as a valuable marker for the management of breast cancer. Methods: The analytical and clinical performance of the Access' BR Monitor Immunoassay System (Beckman Coulter) was evaluated at five different European sites and compared with a reference system, defined as CA15-3 on the Elecsys(R) System (Roche Diagnostics). Results: Total imprecision (%CV) of the BR Monitor ranged between 5.5% and 11.7%, and inter-laboratory reproducibility between 3.4% and 5.1%. Linearity upon dilution showed a mean recovery of 98.5% (SD+/-9.1%). Endogenous interferents had no influence on BR Monitor levels (mean recoveries: hemoglobin 112%, bilirubin 111%, triglycerides 108%). There was no high-dose hook effect up to 13,540 kU/L. Clinical performance investigated in sera from 1811 individuals showed a general correlation between the Access BR Monitor and Elecsys CA15-3 (R = 0.797), with a slope of 1.383. CA15-3 serum levels, as measured by the BR Monitor, were low in healthy individuals (n = 267, median = 11.9 kU/L, 95th percentile = 23.5 kU/L), higher in individuals with various benign diseases (n = 549, medians = 11.3-15.6 kU/L, 95th percentiles = 21.6-54.6 kU/L) and even higher in individuals suffering from various cancers (n = 995, medians = 11.2-22.8 kU/L, 95th percentiles = 30.0-429.7 kU/L). Best diagnostic accuracy for cancer detection against the relevant benign control group by the BR Monitor was found for locoregional and metastatic breast cancer, as well as for ovarian cancer {[}area under the curve (AUC) 0.619, 0.897 and 0.774]. Results for the reference CA15-3 assay were comparable (AUC 0.611, 0.887 and 0.818). Conclusions: The Access BR Monitor provides accurate methodological characteristics and demonstrates an analytical and clinical correlation with Elecsys CA15-3. Best diagnostic accuracy for the BR Monitor was found in breast and ovarian cancer. Our results also suggest a clinical value of the BR Monitor in other cancers.

Dokument bearbeiten Dokument bearbeiten