Abstract
We describe the modification of an existing method of ligament strain measurement at the knee joint in detail. At ten fresh joint specimens we used that technique where strain gauges are attached to the ligamentous insertions and origins. We both improved the preparation of the attachment site and the application of the strain gauges. In a special apparatus the specimens were moved from 0degrees extension to 100degrees flexion while simulating muscle strength and axial force. Testing was performed at the posterior cruciate ligament with both intact and transsected anterior cruciate ligament. In contrast to other existing techniques it does not affect the motion of the joint or the integrity and the function of the ligaments. Unlike the original description of that method we could register a loading behaviour of the posterior cruciate ligament that is similar to those reported in the literature.