Vogeser, Michael; Weigand, Michael; Fraunberger, Peter; Fischer, Heike und Cremer, Peter (2000): Evaluation of the ADVIA (R) Centaur (TM) TSH-3 assay. In: Clinical Chemistry and Laboratory Medicine, Bd. 38, Nr. 4: S. 331-334 [PDF, 160kB]

[thumbnail of cclm.2000.048.pdf]
Vorschau
Download (160kB)

Abstract

An analytical evaluation of the thyroid stimulating hormone (TSH-3) assay on the Sayer ADVIA(R) Centaur(TM) immunoassay system was performed. General analytical requirements (linearity, resistance to typical interferences, absence of a carry-over effect) were fulfilled and reproducibility was satisfactory. Inter-assay coefficient of variation (CV) of a human serum pool with a concentration of 0.014 mU/l was 22.3%; at concentrations between 0.26 and 83 mU/l CV was below 6%. Method comparison study demonstrated close agreement of TSH results compared to those obtained with the Roche Elecsys(R) 2010 TSH assay (ADVIA Centaur = 1.08 x Elecsys - 0.18 mU/l; r = 0.987; n = 324). Handling and practicability of the ADVIA Centaur system proved to be convenient with a very high sample throughput. We conclude that the ADVIA Centaur TSH-3 assay meets requirements for clinical use.