Abstract
An analytical evaluation of the thyroid stimulating hormone (TSH-3) assay on the Sayer ADVIA(R) Centaur(TM) immunoassay system was performed. General analytical requirements (linearity, resistance to typical interferences, absence of a carry-over effect) were fulfilled and reproducibility was satisfactory. Inter-assay coefficient of variation (CV) of a human serum pool with a concentration of 0.014 mU/l was 22.3%; at concentrations between 0.26 and 83 mU/l CV was below 6%. Method comparison study demonstrated close agreement of TSH results compared to those obtained with the Roche Elecsys(R) 2010 TSH assay (ADVIA Centaur = 1.08 x Elecsys - 0.18 mU/l; r = 0.987; n = 324). Handling and practicability of the ADVIA Centaur system proved to be convenient with a very high sample throughput. We conclude that the ADVIA Centaur TSH-3 assay meets requirements for clinical use.
Item Type: | Journal article |
---|---|
Form of publication: | Publisher's Version |
Faculties: | Medicine |
Subjects: | 600 Technology > 610 Medicine and health |
URN: | urn:nbn:de:bvb:19-epub-17888-1 |
ISSN: | 1434-6621 |
Alliance/National Licence: | This publication is with permission of the rights owner freely accessible due to an Alliance licence and a national licence (funded by the DFG, German Research Foundation) respectively. |
Language: | English |
Item ID: | 17888 |
Date Deposited: | 02. Jan 2014, 10:34 |
Last Modified: | 04. Nov 2020, 12:59 |