Abstract
Objectives: The aim of this study was to assess the impact of the advanced technology of the new ExAblate 2100 system (Insightec Ltd, Haifa, Israel) for magnetic resonance imaging (MRI)-guided focused ultrasound surgery on treatment outcomes in patients with symptomatic uterine fibroids, as measured by the nonperfused volume ratio. Materials and Methods: This is a retrospective analysis of 115 women (mean age, 42 years; range, 27-54 years) with symptomatic fibroids who consecutively underwent MRI-guided focused ultrasound treatment in a single center with the new generation ExAblate 2100 system from November 2010 to June 2011. Mean +/- SD total volume and number of treated fibroids (per patient) were 89 +/- 94 cm(3) and 2.2 +/- 1.7, respectively. Patient baseline characteristics were analyzed regarding their impact on the resulting nonperfused volume ratio. Results: Magnetic resonance imaging-guided focused ultrasound treatment was technically successful in 115 of 123 patients (93.5%). In 8 patients, treatment was not possible because of bowel loops in the beam pathway that could not be mitigated (n = 6), patient movement (n = 1), and system malfunction (n = 1). Mean nonperfused volume ratio was 88% +/- 15% (range, 38%-100%). Mean applied energy level was 5400 +/- 1200 J, and mean number of sonications was 74 +/- 27. No major complications occurred. Two cases of first-degree skin burn resolved within 1 week after the intervention. Of the baseline characteristics analyzed, only the planned treatment volume had a statistically significant impact on nonperfused volume ratio. Conclusions: With technological advancement, the outcome of MRI-guided focused ultrasound treatment in terms of the nonperfused volume ratio can be enhanced with a high safety profile, markedly exceeding results reported in previous clinical trials.
Dokumententyp: | Zeitschriftenartikel |
---|---|
Publikationsform: | Publisher's Version |
Keywords: | MRI-guided focused ultrasound; leiomyoma; uterine fibroid; ablation; nonperfused volume |
Fakultät: | Medizin |
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
URN: | urn:nbn:de:bvb:19-epub-24130-2 |
ISSN: | 0020-9996 |
Ort: | 530 WALNUT ST, PHILADELPHIA, PA 19106-3621 USA |
Allianz-/Nationallizenz: | Dieser Beitrag ist mit Zustimmung des Rechteinhabers aufgrund einer (DFG-geförderten) Allianz- bzw. Nationallizenz frei zugänglich. |
Sprache: | Englisch |
Dokumenten ID: | 24130 |
Datum der Veröffentlichung auf Open Access LMU: | 16. Mrz. 2015, 09:04 |
Letzte Änderungen: | 04. Nov. 2020, 13:05 |