Abstract
The wear behaviour of total knee arthroplasty (TKA) is dominated by two wear mechanisms: the abrasive wear and the delamination of the gliding components, where the second is strongly linked to aging processes and stress concentration in the material. The addition of vitamin E to the bulk material is a potential way to reduce the aging processes. This study evaluates the wear behaviour and delamination susceptibility of the gliding components of a vitamin E blended, ultra-high molecular weight polyethylene (UHMWPE) cruciate retaining (CR) total knee arthroplasty. Daily activities such as level walking, ascending and descending stairs, bending of the knee, and sitting and rising from a chair were simulated with a data set received from an instrumented knee prosthesis. After 5 million test cycles no structural failure of the gliding components was observed. The wear rate was with 5.62 +/- 0.53 mg/million cycles falling within the limit of previous reports for established wear test methods.
Item Type: | Journal article |
---|---|
Form of publication: | Publisher's Version |
Faculties: | Medicine |
Subjects: | 600 Technology > 610 Medicine and health |
URN: | urn:nbn:de:bvb:19-epub-24314-4 |
ISSN: | 2314-6133 |
Language: | English |
Item ID: | 24314 |
Date Deposited: | 27. Mar 2015, 10:46 |
Last Modified: | 04. Nov 2020, 13:05 |