Maier, Barbara und Vogeser, Michael (April 2013): Target analyte quantification by isotope dilution LC-MS/MS directly referring to internal standard concentrations-validation for serum cortisol measurement. In: Clinical chemistry and laboratory medicine : CCLM / FESCC, Bd. 51, Nr. 4: S. 833-837 [PDF, 372kB]

[thumbnail of Maier_Vogeser_Target.pdf]
Vorschau
Download (372kB)

Abstract

Isotope dilution LC-MS/MS methods used in the clinical laboratory typically involve multi-point external calibration in each analytical series. Our aim was to test the hypothesis that determination of target analyte concentrations directly derived from the relation of the target analyte peak area to the peak area of a corresponding stable isotope labelled internal standard compound [direct isotope dilution analysis (DIDA)] may be not inferior to conventional external calibration with respect to accuracy and reproducibility.