Abstract
Introduction. Overlooked polymethylmethacrylate after unicondylar knee arthroplasty can be a potential problem, since this might influence the generated wear particle size and morphology. The aim of this study was the analysis of polyethylene wear in a knee wear simulator for changes in size, morphology, and particle number after the addition of third-bodies. Material and Methods. Fixed bearing unicondylar knee prostheses (UKA) were tested in a knee simulator for 5.0 million cycles. Following bone particles were added for 1.5million cycles, followed by 1.5million cycles with PMMA particles. A particle analysis by scanning electron microscopy of the lubricant after the cycles was performed. Size and morphology of the generated wear were characterized. Further, the number of particles per 1 million cycles was calculated for each group. Results. The particles of all groups were similar in size and shape. The number of particles in the PMMA group showed 10-fold higher values than in the bone and control group (PMMA: 10.251 x 10(12); bone: 1.145x10(12); control: 1.804x10(12)). Conclusion. The addition of bone or PMMA particles in terms of a third-body wear results in no change of particle size and morphology. PMMA third-bodies generated tenfold elevated particle numbers. This could favor an early aseptic loosening.
| Dokumententyp: | Zeitschriftenartikel |
|---|---|
| Fakultät: | Medizin |
| Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
| URN: | urn:nbn:de:bvb:19-epub-34585-3 |
| ISSN: | 2314-6133 |
| Sprache: | Englisch |
| Dokumenten ID: | 34585 |
| Datum der Veröffentlichung auf Open Access LMU: | 15. Feb. 2017 16:04 |
| Letzte Änderungen: | 04. Nov. 2020 13:13 |