Logo Logo
Hilfe
Hilfe
Switch Language to English

Freudenthaler, Volker; Seefeldner, Meinhard; Groß, Silke und Wandinger, Ulla (2016): Accuracy of linear depolarisation ratios in clean air ranges measured with POLIS-6 at 355 and 532 nm. In: EPJ Web of Conferences, Bd. 119, 25013 [PDF, 1MB]

[thumbnail of 10.1051_epjconf_201611925013.pdf]
Vorschau

Download (1MB)

Abstract

Linear depolarization ratios in clean air ranges were measured with POLIS-6 at 355 and 532 nm The mean deviation from the theoretical values, including the rotational Raman lines within the filter bandwidths, amounts to 0.0005 at 355 nm and to 0.0012 at 532 nm The mean uncertainty of the measured linear depolarization ratio of clean air is about 0.0005 at 355 nm and about 0.0006 at 532 nm

Dokument bearbeiten Dokument bearbeiten