Schuh, S.; Kaestle, R.; Sattler, E. und Welzel, J. (2016): Comparison of different optical coherence tomography devices for diagnosis of non-melanoma skin cancer. In: Skin Research and Technology, Bd. 22, Nr. 4: S. 395-405

Volltext auf 'Open Access LMU' nicht verfügbar.

Abstract

PurposeTo compare the diagnostic imaging ability of three different optical coherence tomography (OCT) devices in non-melanoma skin cancer (NMSC). MethodsThirty actinic keratoses (AKs) and 27 basal cell carcinomas (BCCs) of 29 patients were examined with three different OCT devices, VivoSight((R)), Callisto((R)) and Skintell((R)). ResultsComplete data sets were available for 16 BCCs and 10 AKs of 18 patients. All OCT devices were able to discriminate BCCs and AKs significantly from perilesional normal skin due to lower signal intensities as well as a thicker stratum corneum and epidermis in AKs. A significant decrease in the signal intensity and thickness of all skin layers was noted with Skintell((R)) in contrast to VivoSight((R)) and Callisto((R)). OCT comparisons revealed only slight differences between VivoSight((R)) and Callisto((R)). Regarding BCC tumor thickness VivoSight((R)) and Callisto((R)) correlated well, histology did not correlate with the three OCT devices, whereas Skintell((R)) showed no correlation with VivoSight((R)), Callisto((R)) or histology. ConclusionAll tested OCT devices could identify BCCs and AKs objectively through standardized measurement of signal intensity and skin layer thickness. Due to their technical specifications (resolution, penetration depth), each of the OCT systems offers additional and special information on NMSC.