Abstract
Purpose: To provide an experimental basis for spectral optimization of iodine-enhanced CT by a quantitative analysis of image quality and radiation dose characteristics consistently measured for a large variety of scan settings at an anthropomorphic phantom. Methods: CT imaging and thermoluminescent dosimetry were performed at an anthropomorphic whole-body phantom with iodine inserts for different tube voltages (U, 70-140 kV) and current-time products (Q, 60-300 mAs). For all U-Q combinations, the iodine contrast (C), the noise level (N) and, from these, the contrast-to-noise ratio (CNR) of reconstructed CT images were determined and parameterized as a function of U, Q or the measured absorbed dose (D). Finally, two characteristic curves were derived that give the relative increase of CNR at constant D and the relative decrease of D at constant CNR when lowering U. Results: Lowering U affects the measured CNR only slightly but markedly reduces D. For example, reducing U from 120 kV to 70 kV increases the CNR at constant D by a factor of nearly 1.8 or, alternatively, reduces D at constant CNR by a factor of nearly 5. Conclusion: Spectral optimization by lowering U is an effective approach to attain the necessary CNR for a specific diagnostic task at hand while at the same time reducing radiation exposure as far as practically achievable. The characteristic curves derived in this study from extensive measurements at a reference 'person' can support CT users in an easy-to-use manner to select an appropriate voltage for various clinical scenarios. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Dokumententyp: | Zeitschriftenartikel |
---|---|
Fakultät: | Medizin |
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
ISSN: | 1120-1797 |
Sprache: | Englisch |
Dokumenten ID: | 44153 |
Datum der Veröffentlichung auf Open Access LMU: | 27. Apr. 2018, 08:05 |
Letzte Änderungen: | 04. Nov. 2020, 13:19 |