Abstract
Fluorescence diagnosis may be used to improve the safety and reliability of stereotactic brain tumor biopsies using biopsy needles with integrated fiber optics. Based on 5-aminolevulinic-acid-induced protoporphyrin IX (PpIX) fluorescence, vital tumor tissue can be localized in vivo during the excision procedure to reduce the number of necessary samples for a reliable diagnosis. In this study, the practical suitability of two different PpIX excitation wavelengths (405nm, 633nm) was investigated on optical phantoms. Violet excitation at 405nm provides a 50-fold higher sensitivity for the bulk tumor;this factor increases up to 100 with decreasing fluorescent volume as shown by ray tracing simulations. Red excitation at 633nm, however, is noticeably superior with regard to blood layers obscuring the fluorescence. Experimental results on the signal attenuation through blood layers of well-defined thicknesses could be confirmed by ray tracing simulations. Typical interstitial fiber probe measurements were mimicked on agarose-gel phantoms. Even in direct contact, blood layers of 20-40 mu m between probe and tissue must be expected, obscuring 405-nm-excited PpIX fluorescence almost completely, but reducing the 633-nm-excited signal only by 25.5%. Thus, 633nm seems to be the wavelength of choice for PpIX-assisted detection of high-grade gliomas in stereotactic biopsy. [GRAPHICS] . PpIX signal attenuation through clinically relevant blood layers for 405 nm (violet) and 633 nm (red) excitation.
Dokumententyp: | Zeitschriftenartikel |
---|---|
Fakultät: | Medizin |
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
ISSN: | 1864-063X |
Sprache: | Englisch |
Dokumenten ID: | 44242 |
Datum der Veröffentlichung auf Open Access LMU: | 27. Apr. 2018, 08:05 |
Letzte Änderungen: | 04. Nov. 2020, 13:20 |