Abstract
Diagnosis and treatment of many lung diseases like cystic fibrosis (CF) or chronic obstructive pulmonary disease (COPD) could benefit from 3D ventilation information. Applying two EIT systems concurrently is a simple approach without specialized hardware that allows monitoring of regional changes of ventilation distribution inside the thorax at different planes with the high temporal resolution much valued in common single plane EIT. Effects of two simultaneously operated EIT devices on one subject were investigated to monitor rapid processes inside the thorax with a multi-plane approach. Results obtained by simulations with a virtual phantom and measurements with a phantom tank reveal that the distance of electrode planes has an important influence on the signal quality. Band-pass filters adapted according to the distance of the planes, can be used to reduce the crosstalk of the concurrent EIT systems. Besides simulations and phantom tank experiments measurements were also taken from a lung healthy volunteer to demonstrate the operation under realistic conditions. Reconstructed images indicate that it is possible to simultaneously visualize regional ventilation at different planes if settings of the EIT devices are chosen appropriately.
Dokumententyp: | Zeitschriftenartikel |
---|---|
Fakultät: | Medizin |
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
ISSN: | 0967-3334 |
Sprache: | Englisch |
Dokumenten ID: | 44844 |
Datum der Veröffentlichung auf Open Access LMU: | 27. Apr. 2018, 08:07 |
Letzte Änderungen: | 04. Nov. 2020, 13:21 |