Logo Logo
Hilfe
Hilfe
Switch Language to English

Omidvari, Negar; Cabello, Jörge; Topping, Geoffrey; Schneider, Florian R.; Paul, Stephan; Schwaiger, Markus und Ziegler, Sibylle I. (2017): PET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner. In: Physics in Medicine and Biology, Bd. 62, Nr. 22: S. 8671-8692

Volltext auf 'Open Access LMU' nicht verfügbar.

Abstract

MADPET4 is the first small animal PET insert with two layers of individually read out crystals in combination with silicon photomultiplier technology. It has a novel detector arrangement, in which all crystals face the center of field of view transaxially. In this work, the PET performance of MADPET4 was evaluated and compared to other preclinical PET scanners using the NEMA NU 4 measurements, followed by imaging a mouse-size hot-rod resolution phantom and two in vivo simultaneous PET/MRI scans in a 7 T MRI scanner. The insert had a peak sensitivity of 0.49%, using an energy threshold of 350 keV. A uniform transaxial resolution was obtained up to 15 mm radial offset from the axial center, using filtered back-projection with single-slice rebinning. The measured average radial and tangential resolutions (FWHM) were 1.38 mm and 1.39 mm, respectively. The 1.2 mm rods were separable in the hot-rod phantom using an iterative image reconstruction algorithm. The scatter fraction was 7.3% and peak noise equivalent count rate was 15.5 kcps at 65.1 MBq of activity. The FDG uptake in a mouse heart and brain were visible in the two in vivo simultaneous PET/MRI scans without applying image corrections. In conclusion, the insert demonstrated a good overall performance and can be used for small animal multi-modal research applications.

Dokument bearbeiten Dokument bearbeiten