Schwahofer, Andrea; Feist, Harald; Georg, Holger; Häring, Peter and Schlegel, Wolfgang (2017): Experimental determination of the photon-energy dependent dose-to-water response of TLD600 and TLD700 (LiF:Mg,Ti) thermoluminescence detectors. In: Zeitschrift für Medizinische Physik, Vol. 27, No. 1: pp. 13-20

Full text not available from 'Open Access LMU'.

Abstract

The aim of this study has been the experimental determination of the energy dependent dose-to-water response of TLD600 and TLD700 thermoluminescent detectors (Harshaw) in X-ray beams with mean photon energies from about 20 to 200 keV in comparison with 6 Co gamma rays and 6 MV X-rays. Experiments were carried out in collaboration with the German secondary standard laboratory PTW Freiburg. The energy dependent relative responses of TLD600 and TLD700 thermoluminescence detectors were determined at radiation qualities between 30 kVp and 280 kVp. The overall uncertainty of the measured values was characterized by standard deviations varying from 1.2 to 3%. The present results agree with previous studies on the energy dependent dose-to-water response of TLD100. As an application example, the results were used to measure doses associated with X-ray imaging in image-guided radiotherapy.