Chen, W.; Bauer, J.; Kurz, C.; Tessonnier, T.; Handrack, J.; Haberer, T.; Debus, J. und Parodi, Katia (2017): A dedicated software application for treatment verification with off-line PET/CT imaging at the Heidelberg Ion Beam Therapy Center. In: Micro-Mini & Nano-Dosimetry & innovative Technologies in Radiation therapy (Mmnd&Itro2016), Bd. 777, 12021

Volltext auf 'Open Access LMU' nicht verfügbar.

Abstract

We present the workflow of the offline-PET based range verification method used at the Heidelberg Ion Beam Therapy Center, detailing the functionalities of an in-house developed software application, SimInterface14, with which range analysis is performed. Moreover, we introduce the design of a decision support system assessing uncertainties and facilitating physicians in decisions making for plan adaptation.