Logo Logo
Hilfe
Hilfe
Switch Language to English

Walter, U.; Noachtar, S. und Hinrichs, H. (2018): Digitale Elektroenzephalographie in der Hirntoddiagnostik. Technische Anforderungen und Ergebnisse einer Umfrage zur Richtlinienkompatibilität digitaler EEG-Systeme bei Anbietern in Deutschland. In: Nervenarzt, Bd. 89, Nr. 2: S. 156-162

Volltext auf 'Open Access LMU' nicht verfügbar.

Abstract

The guidelines of the German Medical Association and the German Society for Clinical Neurophysiology and Functional Imaging (DGKN) require a high procedural and technical standard for electroencephalography (EEG) as an ancillary method for diagnosing the irreversible cessation of brain function (brain death). Nowadays, digital EEG systems are increasingly being applied in hospitals. So far it is unclear to what extent the digital EEG systems currently marketed in Germany meet the guidelines for diagnosing brain death. In the present article, the technical und safety-related requirements for digital EEG systems and the EEG documentation for diagnosing brain death are described in detail. On behalf of the DGKN, the authors sent out a questionnaire to all identified distributors of digital EEG systems in Germany with respect to the following technical demands: repeated recording of the calibration signals during an ongoing EEG recording, repeated recording of all electrode impedances during an ongoing EEG recording, assessability of intrasystem noise and galvanic isolation of measurement earthing from earthing conductor (floating input). For 15 of the identified 20 different digital EEG systems the specifications were provided by the distributors (among them all distributors based in Germany). All of these EEG systems are provided with a galvanic isolation (floating input). The internal noise can be tested with all systems;however, some systems do not allow repeated recording of the calibration signals and/or the electrode impedances during an ongoing EEG recording. The majority but not all of the currently available digital EEG systems offered for clinical use are eligible for use in brain death diagnostics as per German guidelines.

Dokument bearbeiten Dokument bearbeiten