Stimmelmayr, Michael; Lang, Anna; Beuer, Florian; Mansour, Sonia; Erdelt, Kurt; Krennmair, Gerald und Güth, Jan-Frederik (2020): Mechanical stability of all-ceramic abutments retained with three different screw materials in two-piece zirconia implants-an in vitro study. In: Clinical Oral Investigations, Bd. 24, Nr. 5: S. 1801-1806

Volltext auf 'Open Access LMU' nicht verfügbar.

Abstract

Objectives: To measure the abutment rotation and fracture load of two-piece zirconia implants screwed with three different abutment screw materials. Material and methods: Thirty-six zirconia implants with 36 zirconia abutments were distributed into 3 test groups: group G connected with gold screws, group T with titanium screws, and group P with peek screws. In the first part of the study, the rotation angle of the abutments was measured. The second part of the study measured the maximum fracture force of adhesively bonded lithium disilicate crowns after artificial aging and fracture modes were reported. Results In group G, the median rotation angle was 8.0 degrees, in group T 11.6 degrees, and in group P 9.5 degrees. After artificial aging, no screw loosening, crown, abutment, or implant fracture occurred. The median fracture force in group G was 250 N, in group T 263 N, and in group P 196 N. Conclusions: Rotation angles and fracture loads of two-piece zirconia implants with gold, titanium, or peek screws showed no significant differences;however, fracture loads showed inferior results for group P.