Abstract
In vitro cell culture experiments are highly important techniques to accelerate drug discovery, conduct safety testing and reduce the need for animal studies. Therefore, automatization may help to enhance the technical precision, reduce external (including operator's) influence on the data and thus improve reliability. Prior to application in scientific studies, validation of automated systems is absolutely necessary. In this study we present the validation of two combined automated pipetting systems to conduct toxicity studies in HaCaT cells consisting of cell seeding, noxious agent exposure and several assays to assess cell survival, apoptosis and interleukin production. After initial validation of pipetting accuracy, we compared homogeneity after automated seeding to plates seeded by expert laboratory technicians. Moreover, automated dispensing of a potentially unstable noxious agent was analyzed in terms of speed and consistency. We found a 2 % technical imprecision for the cell survival assay and 4.5-6 % for the other assays, bioluminescent and ELISA techniques. Thus, we could demonstrate the excellent technical precision of our assays. In a final step, we found that intraday variations, though acceptable, were much larger than technical variations and had to assume an intraday biological variability between different wells of the same experimental group.
Dokumententyp: | Zeitschriftenartikel |
---|---|
Fakultät: | Medizin |
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
ISSN: | 0378-4274 |
Sprache: | Englisch |
Dokumenten ID: | 85946 |
Datum der Veröffentlichung auf Open Access LMU: | 25. Jan. 2022, 09:16 |
Letzte Änderungen: | 25. Jan. 2022, 09:16 |