Abstract
This paper presents a universal point-of-care system for fully automated quantification of human T-cell lymphotropic virus type 1 (HTLV-1) proviral load, including genomic RNA, based on digital reverse RNA transcription and c-DNA amplification by MD LAMP (mediator displacement loop-mediated isothermal amplification). A disposable microfluidic LabDisk with pre-stored reagents performs automated nucleic acid extraction, reaction setup, emulsification, reverse transcription, digital DNA amplification, and quantitative fluorogenic endpoint detection with universal reporter molecules. Automated nucleic acid extraction from a suspension of HTLV-1-infected CD4+ T-lymphocytes (MT-2 cells) yielded 8 +/- 7 viral nucleic acid copies per MT-2 cell, very similar to the manual reference extraction (7 +/- 2 nucleic acid copies). Fully automated sample processing from whole blood spiked with MT-2 cells showed a comparable result of 7 +/- 3 copies per MT-2 cell after a run time of two hours and 10 min.
Dokumententyp: | Zeitschriftenartikel |
---|---|
Fakultät: | Medizin |
Themengebiete: | 600 Technik, Medizin, angewandte Wissenschaften > 610 Medizin und Gesundheit |
Sprache: | Englisch |
Dokumenten ID: | 96873 |
Datum der Veröffentlichung auf Open Access LMU: | 05. Jun. 2023, 15:24 |
Letzte Änderungen: | 17. Okt. 2023, 14:53 |