Logo Logo
Hilfe
Hilfe
Switch Language to English

Freudenthaler, Volker; Seefeldner, Meinhard; Groß, Silke und Wandinger, Ulla (2015): Accuracy of linear depolarisation ratios in clear air ranges measured with POLIS-6 at 355 and 532 nm. 27th International Laser Radar Conference, New York City, USA, 05. - 10. July 2015. [PDF, 457kB]

Abstract

Linear depolarization ratios in clean air ranges were measured with POLIS-6 at 355 and 532 nm. The mean deviation from the theoretical values, including the rotational Raman lines within the filter bandwidths, amounts to 0.0005 at 355 nm and to 0.0012 at 532 nm. The mean uncertainty of the measured linear depolarization ratio of clean air is about 0.0005 at 355 nm and about 0.0006 at 532 nm.

Dokument bearbeiten Dokument bearbeiten